Camlog and Conelog® SCREWS
Camlog
Camlog implants and components are distributed through
Henry Schein®
in the USA.
Fig. 1
Two implant platform diameter dependent final screws are produced for
Esthomic®, Logfit®, UCLA, Ceramic, Universal, Titanium CAD/CAM Base and
Telescope abutments. The screws are not plated or coated with any material
and are silver in color. 3.3, 3.8, and 4.3mm diameter implants utilize
catalog number J4005.1601 (Fig.1, left) as a final screw and 5.0, 6.0mm implant
diameters utilize catalog number J4005.2001 (Fig.1, right). Both screws are serviced with
a 0.050”- 1.25mm hex driver and torqued to 20Ncm.
Fig. 2
Corresponding laboratory screws are anodized with a brown surface coating
and should only be used for laboratory procedures. 3.3, 3.8, and 4.3mm
diameter implants utilize catalog number J4006.1601 (Fig.2, left) as a final screw and
5.0, 6.0mm implant diameters utilize catalog number J4006.2001 (Fig.2, right). Both screws
are serviced with a 0.050”- 1.25mm hex driver.
Conelog®
Fig. 3
Two implant platform diameter dependent final screws are produced for
Esthomic®, Logfit®, UCLA, Ceramic, Universal, Titanium CAD/CAM
Base and Telescope abutments. The screws are not plated or coated with any
material and are silver in color. 3.3, 3.8, and 4.3mm diameter implants
utilize catalog number C4005.1601 (Fig.3, left) as a final screw and 5.0mm implant
diameters utilize catalog number C4005.2001 (Fig3, right). Both screws are serviced with
a 0.050”- 1.25mm hex driver and torqued to 20Ncm.
Fig. 4
Corresponding laboratory screws are anodized with a brown surface coating
and should only be used for laboratory procedures. 3.3, 3.8, and 4.3mm
diameter implants utilize catalog number C4006.1601 (Fig.4, left) as a final screw and
5.0mm implant diameters utilize catalog number C4006.2001 (Fig.4, right). Both screws
are serviced with a 0.050”- 1.25mm hex driver.